DD.3.2.1.2 Status

The status values for N-CREATE which are specific for these SOP Classes are defined as follows:

Table DD.3.2.1.2-1

RT ION MACHINE VERIFICATION SOP CLASS N-CREATE STATUS VALUES

Status Meaning Code
Success Machine Verification successfully created 0000
Failure No such object instance – Referenced RT Plan not found C227
The Referenced Fraction Group Number does not exist in the referenced plan C221
No beams exist within the referenced fraction group C222
SCU already verifying and cannot currently process this request. C223

The status values for N-SET which are specific for these SOP Classes are defined as follows:

Table DD.3.2.1.2-2

RT ION MACHINE VERIFICATION SOP CLASS N-SET STATUS VALUES

Status Meaning Code
Success Machine Verification successfully updated 0000
Failure Referenced Beam Number not found within the referenced Fraction Group C224
Referenced device or accessory not supported C225
Referenced device or accessory not found within the referenced beam C226