C.8.26.6.1 Ophthalmic Patient Clinical Information and Test Lens Parameters Macro

Table C.8.26.6-2 describes the attributes for the Ophthalmic Patient Clinical and Test Lens Parameters Macro.

Table C.8.26.6-2OPHTHALMIC PATIENT CLINICAL INFORMATION AND TEST LENS PARAMETERS MACRO ATTRIBUTES

Attribute Name Tag Type Attribute Description
Refractive Parameters Used on Patient Sequence (0024,0112) 2 Refractive parameters used when performing visual field test. Zero or one Item shall be included in this sequence.
>Sphere Lens Power (0022,0007) 1 Sphere value in diopters.
>Cylinder Lens Power (0022,0008) 1 Cylinder value in diopters.
>Cylinder Axis (0022,0009) 1 Axis value in degrees.
Pupil Size (0046,0044) 2 The horizontal diameter measurement of the pupil, in mm. 
Pupil Dilated (0022,000D) 2 The patient’s pupils were pharmacologically dilated for this acquisition. Enumerated Values: YES NO If this tag is empty, no information is available.
Intra Ocular Pressure (0022,000B) 3 Value of intraocular pressure in mmHg.
Visual Acuity Measurement Sequence (0024,0110) 3 Measurements of a patient’s visual acuity. Only a single Item is permitted in this sequence.
>Include ‘Visual Acuity Measurement Macro' Table C.8.25.12-2