C.8.21.6 Breast View Module

Table C.8.21.6-1 contains IOD Attributes that describe the view of a Breast Tomosynthesis Image.

Table C.8.21.6-1 BREAST VIEW MODULE ATTRIBUTES

Attribute Name Tag Type Attribute Description
View Code Sequence (0054,0220) 1 Sequence that describes the view of the patient anatomy in this image. Only a single Item shall be included in this sequence.
>Include ‘Code Sequence Macro’ Table 8.8-1. Defined CID 4014.
>View Modifier Code Sequence (0054,0222) 2 Sequence that provides modifiers for the view of the patient anatomy. Zero or more Items shall be included in this sequence.
>>Include ‘Code Sequence Macro’ Table 8.8-1. Defined CID 4015.
Breast Implant Present (0028,1300) 1C Whether or not the imaged breast contains a breast implant regardless of the visibility of a breast implant in the Pixel Data. Enumerated Values: YES NO Required if Modality (0008,0060) is MG. May be present otherwise. Note: The value is expected to be YES for all images acquired on a breast that contains a breast implant, even when a breast implant is displaced during image acquisition.
Partial View (0028,1350) 3 Indicates whether this image is a partial view, that is a subset of a single view of the breast. Enumerated Values: YES, NO If this Attribute is absent, then the image may or may not be a partial view. Note: This may occur when a breast is larger than the active area of the detector. If this Attribute is present, its value shall be NO if there is a View Modifier Code Sequence (0054,0222) Item of value (R-102D6, SRT, “Magnification”) or (R-102D7, SRT, “Spot Compression”).
Partial View Code Sequence (0028,1352) 1C Sequence that describes the portion or section of the breast captured in a partial view image. One or two Items shall be included in this sequence. See C.8.11.7.1.3. Required if Partial View (0028,1350) is present with a value of YES.
>Include ‘Code Sequence Macro’ (Table 8.8-1). Defined CID 4005.