C.8.17.3 Ophthalmic Photographic Parameters Module

This Module describes equipment used to create original images.

Table C.8.17.3-1OPHTHALMIC PHOTOGRAPHIC PARAMETERS MODULE ATTRIBUTES

Attribute Name Tag Type Attribute Description
Acquisition Device Type Code Sequence (0022,0015) 1 Describes the type of acquisition device. Only a single item shall be included in this sequence.
>Include ‘Code Sequence Macro’ Table 8.8.1 Baseline CID 4202
Illumination Type Code Sequence (0022,0016) 2 Coded value for illumination. Zero or one item shall be included in this sequence.
>Include ‘Code Sequence Macro’ Table 8.8.1 Baseline CID 4203
Light Path Filter Type Stack Code Sequence (0022,0017) 2 Filters used in the light source path. Zero or more items shall be included in this sequence.
>Include ‘Code Sequence Macro’ Table 8.8.1 Baseline CID 4204
Light Path Filter Pass-Through Wavelength (0022,0001) 3 Nominal pass-through wavelength of light path filter in nm
Light Path Filter Pass Band (0022,0002) 3 Pass band of light path filter in nm. This Attribute has two Values. The first is the shorter and the second the longer wavelength relative to the peak. The values are for the - 3dB nominal (1/2 of peak) pass through intensity. One of the two Values may be zero length, in which case it is a cutoff filter.
Image Path Filter Type Stack Code Sequence (0022,0018) 2 Describes stack of filters used in image path. Zero or more items shall be included in this sequence.
>Include ‘Code Sequence Macro’ Table 8.8.1 Baseline CID 4204
Image Path Filter Pass-Through Wavelength (0022,0003) 3 Nominal pass-through wavelength of image path filter in nm
Image Path Filter Pass Band (0022,0004) 3 Pass band of image path filter in nm. This Attribute has two Values. The first is the shorter and the second the longer wavelength relative to the peak. The values are for the - 3dB nominal (1/2 of peak) pass through intensity. One of the two Values may be zero length, in which case it is a cutoff filter
Lenses Code Sequence (0022,0019) 2 Lenses that were used during the image acquisition. Zero or more items shall be included in this sequence.
>Include ‘Code Sequence Macro’ Table 8.8.1 Baseline CID 4205
Detector Type (0018,7004) 2 Type of detector used for creating this image. Defined terms: CCD = Charge Coupled Devices CMOS = Complementary Metal Oxide Semiconductor
Channel Description Code Sequence (0022,001A) 1C Describes the light color used for each channel to generate the image. Required if this differs from the natural interpretation. Note: Interpretation and representation of RGB images rely on the assumption that the red channel really contains the red wavelength range of illumination light, the blue channel the blue wavelength range, etc. Some modalities use the RGB Photometric Interpretation as a container representing 3 channels of any illumination wavelength. Shall have the same number of items as the Value of Samples per Pixel Used (0028,0003) if present, or otherwise the value of Samples per Pixel (0028,0002). The channels shall be described in the order in which the channels are encoded.
>Include ‘Code Sequence Macro’ Table 8.8.1 Baseline CID 4206
Camera Angle of View (0022,001E) 3 The aperture angle of the camera, in degrees