A.7.4 US Multi-Frame Image IOD Module Table

Table A.7-1 US MULTI-FRAME IMAGE IOD MODULES

IE Module Reference Usage
Patient Patient C.7.1.1 M
Clinical Trial Subject C.7.1.3 U
Study General Study C.7.2.1 M
Patient Study C.7.2.2 U
Clinical Trial Study C.7.2.3 U
Series General Series C.7.3.1 M
Clinical Trial Series C.7.3.2 U
Frame of Frame of Reference C.7.4.1 U
Reference Synchronization C.7.4.2 C - Required if Modality (0008,0060) = IVUS. May be present otherwise.
Equipment General Equipment C.7.5.1 M
Image General Image C.7.6.1 M
Image Pixel C.7.6.3 M
Contrast/bolus C.7.6.4 C - Required if contrast media was used in this image.
Cine C.7.6.5 M
Multi-frame C.7.6.6 M
Frame Pointers C.7.6.9 U
Palette Color Lookup Table C.7.9 C - Required if Photometric Interpretation (0028,0004) has a value of PALETTE COLOR
Device C.7.6.12 U
Specimen C.7.6.22 U
US Region Calibration C.8.5.5 U
US Image C.8.5.6 M
VOI LUT C.11.2 U
ICC Profile C.11.15 U
SOP Common C.12.1 M
Frame Extraction C.12.3 C - Required if the SOP Instance was created in response to a Frame-Level retrieve request

Notes: 1. For the purpose of conveying ultrasound protocol data management information it is recommended that the Performed Protocol Code Sequence (0040,0260) be assigned the code value(s) of the performed ultrasound protocol, if any. The Baseline Context Group for these code values is Context ID 12001 (defined in PS 3.16).

2. The US Frame of Reference Module was previously included in this IOD, but has been retired. See PS 3.3 2003.

A.7.4.1 Mutually Exclusive IEs

Note: A Curve IE was previously included in this IOD that was mutually exclusive with the Image IE, but has been retired. See PS 3.3 2004.