A.56.1.4 Enhanced PET Image Functional Group Macros

Table A.56-2 specifies the use of the Functional Group macros used in the Multi-frame Functional Group Module for the Enhanced PET Image IOD.

Table A.56-2ENHANCED PET IMAGE FUNCTIONAL GROUP MACROS

Function Group Macro Section Usage
Pixel Measures C.7.6.16.2.1 M
Frame Content C.7.6.16.2.2 M - May not be used as a Shared Functional Group.
Plane Position (Patient) C.7.6.16.2.3 M
Plane Orientation (Patient) C.7.6.16.2.4 M
Referenced Image C.7.6.16.2.5 C - Required if the image or frame has been planned on another image or frame, may be present otherwise.
Derivation Image C.7.6.16.2.6 C - Required if the image or frame has been derived from another SOP Instance.
Frame Anatomy C.7.6.16.2.8 M
Pixel Value Transformation C.7.6.16.2.9 M
Frame VOI LUT C.7.6.16.2.10 M
Real World Value Mapping C.7.6.16.2.11 M - The Defined Context ID for Measurement Units Code Sequence shall be CID 84.
Cardiac Synchronization C.7.6.16.2.7 C - Required if Cardiac Synchronization Technique (0018,9037) equals other than NONE. May be present otherwise.
Respiratory Synchronization C.7.6.16.2.17 C - Required if Respiratory Motion Compensation Technique (0018,9170) equals other than NONE. May be present otherwise.
Radiopharmaceutical Usage C.7.6.16.2.19 M
Patient Physiological State C.7.6.16.2.20 C - Required for cardiac rest and stress images.
PET Frame Type C.8.22.5.1 M
PET Frame Acquisition C.8.22.5.2 C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL. May be present otherwise.
PET Detector Motion Details C.8.22.5.3 C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL and Type of Detector Motion (0054,0202) is not equal to STATIONARY.
PET Position C.8.22.5.4 C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL. May be present otherwise.
PET Frame Correction Factors C.8.22.5.5 C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL. May be present otherwise.
PET Reconstruction C.8.22.5.6 C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL. May be present otherwise.
PET Table Dynamics C.8.22.5.7 C - Required if Image Type (0008,0008) Value 1 equals ORIGINAL and Table Motion (0018,1134) is equal to DYNAMIC.