A.55.4 Breast Tomosynthesis Image Functional Group Macros

Table A.55-2 specifies the use of the Functional Group macros used in the Multi-frame Functional Groups Module for the Breast Tomosynthesis Image IOD.

Table A.55-2BREAST TOMOSYNTHESIS IMAGE FUNCTIONAL GROUP MACROS

Functional Group Macro Section Usage
Pixel Measures C.7.6.16.2.1 M
Frame Content C.7.6.16.2.2 M - May not be used as a Shared Functional Group.
Plane Position (Patient) C.7.6.16.2.3 M
Plane Orientation (Patient) C.7.6.16.2.4 M
Referenced Image C.7.6.16.2.5 U
Derivation Image C.7.6.16.2.6 C - Required if the Image Type (0008,0008) Value 1 equals DERIVED.
Frame Anatomy C.7.6.16.2.8 M
Identity Pixel Value Transformation C.7.6.16.2.9b M
Frame VOI LUT With LUT C.7.6.16.2.10b M
Real World Value Mapping C.7.6.16.2.11 U
Contrast/Bolus Usage C.7.6.16.2.12 C - Required if the Enhanced Contrast/Bolus Module is present.
X-Ray 3D Frame Type C.8.21.5.1 M - May not be used as a Shared Functional Group.

A.55.4.1 Breast Tomosynthesis Image Functional Group Macros Content Constraints

A.55.4.1.1 Frame Anatomy Functional Group Macro

The Defined Context ID for the Anatomic Region Sequence (0008,2218) shall be CID 4013.