A.53.3 X-Ray 3D Angiographic Image IOD Image Module Table

Table A.53-1X-RAY 3D ANGIOGRAPHIC IMAGE IOD MODULES

IE Module Reference Usage
Patient Patient C.7.1.1 M
Clinical Trial Subject C.7.1.3 U
Study General Study C.7.2.1 M
Patient Study C.7.2.2 U
Clinical Trial Study C.7.2.3 U
Series General Series C.7.3.1 M
Clinical Trial Series C.7.3.2 U
Enhanced Series C.7.3.3 M
Frame of Reference Frame of Reference C.7.4.1 M
Equipment General Equipment C.7.5.1 M
Enhanced General Equipment C.7.5.2 M
Image Image Pixel C.7.6.3 M
Enhanced Contrast/Bolus C.7.6.4b C - Required if contrast media was applied.
Device C.7.6.12 U
Intervention C.7.6.13 U
Acquisition Context C.7.6.14 M
Multi-frame Functional Groups C.7.6.16 M
Multi-frame Dimension C.7.6.17 U
Cardiac Synchronization C.7.6.18.1 C - Required if cardiac synchronization was applied.
Respiratory Synchronization C.7.6.18.2 C - Required if respiratory synchronization was applied.
Patient Orientation C.7.6.20 U
Image - Equipment Coordinate Relationship C.7.6.21 U
Specimen C.7.6.22 U
X-Ray 3D Image C.8.21.1 M
X-Ray 3D Angiographic Image Contributing Sources C.8.21.2.1 U
X-Ray 3D Angiographic Acquisition C.8.21.3.2 U
X-Ray 3D Reconstruction C.8.21.4 U
SOP Common C.12.1 M
Frame Extraction C.12.3 C - Required if the SOP Instance was created in response to a Frame-Level retrieve request

A.53.3.1 X-Ray 3D Angiographic Image IOD Content Constraints

A.53.3.1.1 Modality Type Attribute

The Modality Type attribute (0008,0060) shall have the value XA.

A.53.3.1.2 Restrictions for Standard Extended SOP Classes

The Overlay Plane Module, VOI LUT Module and Softcopy Presentation LUT Module shall not be used in a Standard Extended SOP Class of the X-Ray 3D Angiographic Image.

Notes: 1. The VOI LUT function is provided by a Frame VOI LUT Functional Group.

2. The Curve Module was previously included in the Image IE for this IOD but has been retired. See PS 3.3 2004.