A.51.3 Segmentation IOD Module Table

Table A.51-1 SEGMENTATION IOD MODULES

IE Module Reference Usage
Patient Patient C.7.1.1 M
Clinical Trial Subject C.7.1.3 U
Study General Study C.7.2.1 M
Patient Study C.7.2.2 U
Clinical Trial Study C.7.2.3 U
Series General Series C.7.3.1 M
Segmentation Series C.8.20.1 M
Clinical Trial Series C.7.3.2 U
Frame of Reference Frame of Reference C.7.4.1 C - Required if Derivation Image Functional Group (C.7.16.2.6) is not present.
Equipment General Equipment C.7.5.1 M
Enhanced General Equipment C.7.5.2 M
Segmentation General Image C.7.6.1 M
Image Pixel C.7.6.3 M
Segmentation Image C.8.20.2 M
Multi-frame Functional Groups C.7.6.16 M
Multi-frame Dimension C.7.6.17 M
Specimen C.7.6.22 U
Common Instance Reference C.12.2 C - Required if Derivation Image Functional Group (C.7.16.2.6) is present.
SOP Common C.12.1 M
Frame Extraction C.12.3 C - Required if the SOP Instance was created in response to a Frame-Level retrieve request

Note: The Specimen Identification Module was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module. See PS 3.3-2008.