A.27 Digital MAMMOGRAPHY X-Ray image information object definition

A.27.1 Digital Mammography X-Ray Image IOD Description

The Digital Mammography X-Ray Image Information Object Definition specifies an image that has been created by a digital mammography projection radiography imaging device.

Note: It meets all of the requirements of the DX IOD in A.26 in addition to those specified in this section.

The Digital Mammography Image IOD is used in two SOP Classes as defined in PS 3.4 Storage Service Class, a SOP Class for storage of images intended for presentation, and a SOP Class for storage of images intended for further processing before presentation. These are distinguished by their SOP Class UID and by the Enumerated Value of the mandatory Attribute in the DX Series Module, Presentation Intent Type (0008,0068).

A.27.2 Digital Mammography X-Ray Image IOD Module Table

Table A.27-1DIGITAL MAMMOGRAPHY X-RAY IMAGE IOD MODULES

IE Module Reference Usage
Patient Patient C.7.1.1 M
Clinical Trial Subject C.7.1.3 U
Study General Study C.7.2.1 M
Patient Study C.7.2.2 U
Clinical Trial Study C.7.2.3 U
Series General Series C.7.3.1 M
Clinical Trial Series C.7.3.2 U
DX Series C.8.11.1 M
Mammography Series C.8.11.6 M
Frame of Reference Frame of Reference C.7.4.1 C - Required if multiple images are obtained without releasing breast compression
Equipment General Equipment C.7.5.1 M
Image General Image C.7.6.1 M
Image Pixel C.7.6.3 M
Contrast/Bolus C.7.6.4 U
Display Shutter C.7.6.11 U
Device C.7.6.12 U
Intervention C.7.6.13 U
Specimen C.7.6.22 U
DX Anatomy Imaged C.8.11.2 M
DX Image C.8.11.3 M
DX Detector C.8.11.4 M
X-Ray Collimator C.8.7.3 U
DX Positioning C.8.11.5 U
X-Ray Tomo Acquisition C.8.7.7 U
X-Ray Acquisition Dose C.8.7.8 U
X-Ray Generation C.8.7.9 U
X-Ray Filtration C.8.7.10 U
X-Ray Grid C.8.7.11 U
Mammography Image C.8.11.7 M
Overlay Plane C.9.2 C - Required if graphic annotation is present - See A.27.3
VOI LUT C.11.2 C - Required if Presentation Intent Type (0008,0068) is FOR PRESENTATION. Shall not be present otherwise.
Image Histogram C.11.5 U
Acquisition Context C.7.6.14 M
SOP Common C.12.1 M

Notes: 1. The Curve Module was previously included in this IOD but has been retired. See PS 3.3 2004.

2. The Specimen Identification Module was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module. See PS 3.3-2008.

A.27.3 Overlay Plane Module

If the Overlay Plane Module is present, any Overlays defined in that Module shall store the overlay data in Overlay Data (60xx,3000), and not any unused high bits in Pixel Data (7FE0,0010).