TID 5202 Echo Section

This is a generic section heading Template for any of the anatomical headings. Measurements within a section heading appear as groups (by image mode, acquisition protocol, and/or protocol stage).

Parameter Name Parameter Usage
$SectionSubject The subject modifier of the section heading container
$MeasType The concept name of the measurement

TID 5202ECHO SECTIONType: Extensible Order: Significant

NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint
1 CONTAINER EV (121070, DCM, ”Findings”) 1 M
2 > HAS CONCEPT MOD CODE EV (G-C0E3, SRT, “Finding Site”) 1 M $SectionSubject
3 > CONTAINS CONTAINER DT (125007, DCM, “Measurement Group”) 1-n M
4 >> HAS CONCEPT MOD CODE EV (G-0373, SRT,”Image Mode” ) 1 U BCID (12224) Ultrasound Image Modes
5 >> HAS CONCEPT MOD CODE DT (125203,DCM,”Acquisition Protocol”) 1 U
6 >> HAS CONCEPT MOD TEXT DT (125203, DCM, “Acquisition Protocol”) 1 U
7 >> HAS ACQ CONTEXT CODE EV (LN, 18139-6, "Stage") 1 U BCID (12002) Ultrasound Protocol Stage Types
8 >> CONTAINS INCLUDE DTID (5203) Echo Measurement 1-n M $Measurement = $MeasType $Method=CID (12227) Echocardiography Measurement Method

Echo Section Descriptions

Rows 4, 5 Type of measurement group. May be grouped by image mode, or acquisition protocol, or some other user or manufacturer designated classification
Row 7 For measurements acquired in a staged protocol, all measurements in a measurement group are acquired at the identified stage.