TID 4105 Chest CAD Descriptors

This template provides qualitative detail for a Single Image Finding or Composite Feature. It is applied to Chest CAD Composite Feature (TID 4102) and Chest CAD Single Image Finding (TID 4104).

TID 4105CHEST CAD DESCRIPTORSType: Non-Extensible Order: Significant

NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint
1 CODE EV (112025, DCM, “Size Descriptor”) 1 U DCID (6118) Size Descriptor
2 CODE EV (112026, DCM, “Width Descriptor”) 1 U DCID (6107) Width Descriptor
3 CODE EV (112015, DCM, “Border shape”) 1 U DCID (6119) Chest Border Shape
4 CODE EV (112007, DCM, “Border definition”) 1 U DCID (6120) Chest Border Definition
5 CODE EV (112014, DCM, “Orientation Descriptor”) 1 U DCID (6121) Chest Orientation Descriptor
6 CODE EV (112009, DCM, “Type of Content”) 1-n U DCID (6122) Chest Content Descriptor
7 CODE EV (112027, DCM, “Opacity Descriptor”) 1 U DCID (6123) Chest Opacity Descriptor
8 CODE EV (112013, DCM, “Location in Chest”) 1 U DCID (6124) Location in Chest
9 CODE EV (G-C171, SRT, “Laterality”) 1 U DCID (244) Laterality
10 CODE EV (112006, DCM, “Distribution Descriptor”) 1-n U DCID (6128) Chest Distribution Descriptor
11 CODE EV (112028, DCM, “Abnormal Distribution of Anatomic Structure”) 1 U DCID (6108) Chest Anatomic Structure Abnormal Distribution
12 CODE EV (112008, DCM, “Site involvement”) 1-n U DCID (6129) Chest Site Involvement
13 CODE EV (G-C197, SRT, “Severity”) 1 U DCID (6130) Severity Descriptor
14 CODE EV (112010, DCM, “Texture Descriptor”) 1 U DCID (6131) Chest Texture Descriptor
15 CODE EV (112030, DCM, “Calcification Descriptor”) 1 U DCID (6132) Chest Calcification Descriptor
16 NUM DCID (6142) Calculated Value 1-n U
17 > HAS CONCEPT MOD CODE EV (121401, DCM, “Derivation”) 1 M DCID (6140) Calculation Methods
18 > INFERRED FROM NUM EV (112032, DCM, “Threshold Attenuation Coefficient”) 1 U UNITS = EV ([hnsf’U], UCUM, “Hounsfield unit”)
19 > INFERRED FROM TEXT EV (112034, DCM, “Calculation Description”) 1 U
20 NUM DCID (6141) Attenuation Coefficient Measurements 1-n U UNITS = EV ([hnsf’U], UCUM, “Hounsfield unit”)