TID 4015 CAD Detections Performed Template

This template gathers two lists of detection algorithms attempted, grouped by success/failure status. Algorithms not attempted are not mentioned in this sub-tree of the Document Root (e.g., TID 4000). This information forms the basis for understanding why a CAD report may produce no (or fewer than anticipated) detection results.

The sub-tree formed by this template is illustrated in PS 3.17 in the Annex on Mammography CAD SR Content Tree Structure.

Parameter Name Parameter Usage
$DetectionCode Coded term or Context Group for Detection Performed

TID 4015 CAD DETECTIONS PERFORMEDType: Non-Extensible Order: Significant

NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint
1 CONTAINER EV(111063, DCM, “Successful Detections”) 1 MC Shall be present only if value of parent is (111222, DCM, “Succeeded”) or (111223, DCM, “Partially Succeeded”)
2 > CONTAINS INCLUDE DTID (4017) CAD Detection Performed 1-n M $DetectionCode = $DetectionCode
3 CONTAINER EV(111025, DCM, “Failed Detections”) 1 MC Shall be present only if value of parent is (111224, DCM, “Failed”) or (111223, DCM, “Partially Succeeded”)
4 > CONTAINS INCLUDE DTID (4017) CAD Detection Performed 1-n M $DetectionCode = $DetectionCode