TID 3907 Vessel Measurements

Contains measurements made on vessel level.

TID 3907Vessel MeasurementsType: Extensible Order: Significant

NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint
1 CONTAINS NUM EV(R-101BB, SRT, “Lumen Diameter Stenosis“) 1 U UNITS=DT(%,UCUM, “%”)
2 CONTAINS NUM EV(R-101BA, SRT, “Lumen Area Stenosis“) 1 U UNITS=DT(%,UCUM, “%”)
3 CONTAINS NUM EV(121206, DCM, ”Distance”) 1-n U UNITS=DT(mm, UCUM, “mm”)
4 > HAS CONCEPT MOD CODE EV(122340, DCM, “Fiducial Feature”) 2 M
5 >> HAS CONCEPT MOD CODE EV(G-C171, SRT, “Laterality”) 1 U DCID(244) Laterality
6 > INCLUDE DTID (320) Image or Spatial Coordinates 1 U
7 CONTAINS NUM EV(G-0364, SRT, "Vessel Lumen Diameter") 1-n U UNITS=DT(mm, UCUM, "mm")
8 > HAS CONCEPT MOD CODE EV (121401, DCM, “Derivation”) 1 U DCID(3488) Min/Max/Mean
9 > HAS CONCEPT MOD NUM EV(122337, DCM, "Relative Position”) 1 U Units=EV(mm, UCUM, "mm")
10 >> HAS CONCEPT MOD CODE EV(122340, DCM, "Fiducial Feature") 1 M DCID(3837) Fiducial Feature

Content Item Descriptions

Rows 3-5 The distance between two identified fiducial features
Rows 7-10 Measurement of vessel diameter made at a position relative to a fiducial feature
Row 9 A positive value indicates a distance in the direction of flow within the vessel