TID 3515 Cardiac Output Measurement by Indicator Dilution

The Cardiac Output Measurement by Indicator Dilution template consists of a CONTAINER containing the measurement and significant parameters of the technical method. If the measurement is based on a DICOM Hemodynamic Waveform SOP Instance, that object may also be referenced.

Note: Fick Cardiac Output is encoded in the Derived Hemodynamic Measurements Template 3560.

TID 3515Cardiac Output Measurement by Indicator DilutionType: Extensible Order: Significant

NL Relation with Parent Value Type Concept Name VM Req Type Condition Value Set Constraint
1 CONTAINER EV (P2-34201, SRT, “Cardiac Output measurement”) 1 M
2 > CONTAINS INCLUDE TID (300) Measurement 1 M $Measurement = EV (8737-9, LN, “Cardiac Output by Indicator Dilution”) $Units = EV (l/min, UCUM, ”l/min”) $Method = DCID (3628) Cardiac Output Methods $WavePurpose = DT (121112, DCM, "Source of measurement”)
3 > HAS ACQ CONTEXT NUM EV (122319, DCM, “Catheter Size”) 1 MC IFF Row 2 indicates a thermal method UNITS = DCID (3510) Catheter Size Units
4 > HAS ACQ CONTEXT NUM EV (122320, DCM, “Injectate Temperature”) 1 MC IFF Row 2 indicates a thermal method UNITS = EV (Cel, UCUM, “°C”)
5 > HAS ACQ CONTEXT NUM EV (122321, DCM, “Injectate Volume”) 1 M UNITS = DT (ml, UCUM, “ml”)
6 > HAS ACQ CONTEXT NUM EV (122322, DCM, “Calibration Factor”) 1 M UNITS = DT (1, UCUM, “no units”)