TID 3501 Hemodynamics Measurement Group

The Hemodynamic Measurement Group template provides a structure for measurements acquired during a single procedure phase in a cardiac catheterization lab.

TID 3501 Hemodynamics Measurement GroupType: Extensible Order: Significant

NL Relation with Parent Value Type Concept Name VM Req Type Condition Value Set Constraint
1 CONTAINER EV (121070, DCM, “Findings”) 1 M
2 > HAS ACQ CONTEXT CODE EV (G-72BB, SRT, “Catheterization Procedure Phase”) 1 M DCID (3651) Hemodynamic Measurement phase
3 > HAS ACQ CONTEXT INCLUDE DTID (3520) Hemodynamic Clinical Context 1 U
4 > HAS ACQ CONTEXT TEXT EV (121124, DCM, "Procedure Action ID") 1 U
5 > CONTAINS INCLUDE DTID (3510) Vital signs 1-n U
6 > CONTAINS INCLUDE DTID (3504) Arterial Pressure Measurement 1-n U
7 > CONTAINS INCLUDE DTID (3505) Atrial Pressure Measurement 1-n U
8 > CONTAINS INCLUDE DTID (3506) Venous Pressure Measurement 1-n U
9 > CONTAINS INCLUDE DTID (3507) Ventricular Pressure Measurement 1-n U
10 > CONTAINS INCLUDE DTID (3508) Gradient Measurement 1-n U
11 > CONTAINS INCLUDE DTID (3509) Blood Velocity Measurements 1-n U
12 > CONTAINS INCLUDE DTID (3515) Cardiac Output Measurement by Indicator Dilution 1-n U
13 > CONTAINS INCLUDE DTID (3516) Blood lab measurements 1-n U
14 > CONTAINS INCLUDE DTID (3560) Derived Hemodynamic Measurements 1-n U
15 > CONTAINS INCLUDE DTID (3714) ECG Lead Measurements 1-n U

Content Item Descriptions

Row 4 Procedure Action ID allows linkage between the hemodynamic measurements recorded in this Template and a procedure step or phase recorded in the Procedure Log, e.g., using TID 3100.