TID 3209 Centerline Wall Motion Template

The Centerline Wall Motion Template consists of a CONTAINER providing measurements of the centerline wall motion.

TID 3209CENTERLINE WALL MOTIONType: Extensible Order: Significant

NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint
1 CONTAINER EV (121070, DCM, “Findings”) 1 M
2 > HAS CONCEPT MOD CODE EV (111004, DCM, “Analysis Performed”) 1 M EV (122449, DCM, “Centerline Wall Motion Analysis”)
3 >> HAS CONCEPT MOD CODE EV (122410, DCM, “Contour Realignment”) 1 M DCID (3458) Contour Realignment
4 > CONTAINS INCLUDE DTID (300) Measurement 100 M $Measurement = EV (122450, DCM, “Normalized Chord Length”) $Unit = DT (%, UCUM, “%”)
5 > CONTAINS NUM EV (122411, DCM, “Threshold Value”) 1 M Values shall be 1, 2 or 3 Units = EV ({sd}, UCUM, “Standard Deviations”)
6 > CONTAINS CONTAINER EV (122451, DCM, “Abnormal Region”) 1-6 U
7 >> CONTAINS CODE EV (F-32050, SRT, “Cardiac Wall Motion”) 1 M DCID (3703) Wall Motion
8 >> CONTAINS CODE EV (R-404F0, SRT, “Circumferential Extent”) 1 U DCID (3460) Circumferential Extent
9 >> CONTAINS NUM EV (122452, DCM, “First Chord of Abnormal Region”) 1 M Unit = DT (1, UCUM, “no unit”)
10 >> CONTAINS NUM EV (122453, DCM, “Last Chord of Abnormal Region”) 1 M Unit = DT (1, UCUM, “no unit”)
11 > CONTAINS CONTAINER EV (122417, DCM, “Regional Abnormal Wall Motion ”) 1-4 U
12 >> HAS CONCEPT MOD CODE EV (G-C03E, SRT, “Finding Site”) 1 M DCID (3461) Regional Extent
13 >> CONTAINS INCLUDE DTID (300) Measurement 1 M $Measurement = EV (122459, DCM, “Territory Region Severity”) $ModType = EV (F-32050, SRT, “Cardiac Wall Motion”) $ModValue = DCID (3703) Wall Motion $Unit = DT ({sd}, UCUM, “Standard Deviations”)
14 >> CONTAINS INCLUDE DTID (300) Measurement 1 M $Measurement = EV (122461, DCM, “Opposite Region Severity”) $ModType = EV (F-32050, SRT, “Cardiac Wall Motion”) $ModValue = DCID (3703) Wall Motion $Unit = DT ({sd}, UCUM, “Standard Deviations”)
15 > CONTAINS IMAGE No purpose of reference 1 U

Content Item Descriptions

Row 4 Normalized lengths of the chords determined between ED and ES contour. The measurement template allows the specification of the statistical properties of the normal population and of the chord measurement relative to the population.
Row 8 If the Circumferential Extent is not specified no limitations to the boundaries for regions are assumed.
Row 11 The Regional Abnormal Wall Motion container holds the information on the severity of the decreased or increased wall motion of the 4 predefined regions as described in [Sheehan, 1986].
Row 12 The name of the region with an abnormal ventricular wall motion as described in [Sheehan, 1986].
Row 13 The severity of the wall motion abnormality expressed in Standard Deviations above or below normal in the territory region as described in [Sheehan, 1986].
Row 14 The severity of the wall motion abnormality expressed in Standard Deviations above or below normal in the opposite region as described in [Sheehan, 1986].
Row 15 Secondary Capture image with centerline analysis result