TID 3206 VA Main Results Template

The VA Main Results Template consists of a CONTAINER with a structure for reporting the main ventricular analysis measurements.

TID 3206 VA MAIN RESULTS Type: Extensible Order: Significant

NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint
1 CONTAINER EV (121070, DCM, “Findings”) 1 M
2 > HAS CONCEPT MOD CODE EV (G-C0E3, SRT, “Finding Site”) 1 M DCID (3462) Chamber Identification
3 > CONTAINS CODE EV (122429, DCM, “Volume Method”) 1 M DCID (3453) Cardiac Volume Methods
4 > CONTAINS NUM EV (122435, DCM, “Regression Volume Exponent”) 1 U Unit = DT (1, UCUM, “no units”)
5 > CONTAINS NUM EV (122431, DCM, “Regression Slope ED”) 1 U Unit = DT (1, UCUM, “ratio”)
6 > CONTAINS NUM EV (122432, DCM, “Regression Offset ED”) 1 U Unit = DT (ml, UCUM, “ml”)
7 > CONTAINS NUM EV (122433, DCM, “Regression Slope ES”) 1 U Unit = DT (1, UCUM, “ratio”)
8 > CONTAINS NUM EV (122434, DCM, “Regression Offset ES”) 1 U Unit = DT (ml, UCUM, “ml”)
9 > CONTAINS INCLUDE DTID (300) Measurement 1 M $Measurement = DCID (3467) Ejection Fraction $Unit = DT (%, UCUM, “%”)
10 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = DCID (3468) ED Volume $Unit = DT (ml, UCUM, “ml”)
11 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = DCID (3469) ES Volume $Unit = DT (ml, UCUM, “ml”)
12 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (F-32120, SRT, “Stroke Volume”) $Unit = DT (ml, UCUM, “ml”)
13 > CONTAINS NUM EV (8867-4, LN, “Heart Rate”) 1 U Unit = DT ({hb}/min, UCUM, “beats/min”)
14 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = DCID (3468) ED Volume $ModType = EV (121425, DCM, “Index”) $ModValue = DCID (3455) Index Methods $Unit = DT (ml/m2, UCUM, “ml/m^ 2”)
15 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = DCID (3468) ED Volume $ModType = EV (121425, DCM, “Index”) $ModValue = EV (29463-7. LN, “Patient Weight”) $Unit = DT (ml/kg, UCUM, “ml/kg”)
16 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = DCID (3469) ES Volume $ModType = EV (121425, DCM, “Index”) $ModValue = DCID (3455) Index Methods $Unit = DT (ml/m2, UCUM, “ml/m^2”)
17 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = DCID (3469) ES Volume $ModType = EV (121425, DCM, “Index”) $ModValue = EV (29463-7. LN, “Patient Weight”) $Unit = DT (ml/kg, UCUM, “ml/kg”)
18 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (F-32120, SRT, “Stroke Volume”) $ModType = EV (121425, DCM, “Index”) $ModValue = DCID (3455) Index Methods $Unit = DT (ml/m2, UCUM, “ml/m^2”)
19 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (F-32120, SRT, “Stroke Volume”) $ModType = EV (121425, DCM, “Index”) $ModValue = EV (29463-7. LN, “Patient Weight”) $Unit = DT (ml/kg, UCUM, “ml/kg”)
20 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (F-32100, SRT, “Cardiac Output”) $Unit = DT (l/min, UCUM, “l/min”)
21 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (F-32110, SRT, “Cardiac Index”) $ModType = EV (121425, DCM, “Index”) $ModValue = DCID (3455) Index Methods $Unit = DT (l/min/m2, UCUM, “l/min/m^2”)
22 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (122445, DCM, “Wall Thickness”) $Unit = DT (mm, UCUM, “mm”)
23 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (122446, DCM, “Wall Volume”) $Unit = DT (ml, UCUM, “ml”)
24 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (122447, DCM, “Wall Mass”) $Unit = DT (g, UCUM, “gram”)
25 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (122447, DCM, “Wall Mass”) $ModType = EV (121425, DCM, “Index”) $ModValue = DCID (3455) Index Methods $Unit = DT (g/m2, UCUM, “gram/m^2”)
26 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (122447, DCM, “Wall Mass”) $ModType = EV (121425, DCM, “Index”) $ModValue = EV (29463-7. LN, “Patient Weight”) $Unit = DT (g/kg, UCUM, “gram/kg”)
27 > CONTAINS INCLUDE DTID (300) Measurement 1 U $Measurement = EV (122448, DCM, “Wall Stress”) $Unit = DT (dyn/cm2, UCUM, “dynes/cm^2”)
28 > CONTAINS IMAGE No purpose of reference 1-n U

Content Item Descriptions

Row 28 Secondary Capture image with ED and/or ES contours