TID 3202 Ventricular Analysis Template

The Ventricular Analysis Template provides a CONTAINER with a structure for reporting the result of the ventricular analysis.

TID 3202VENTRICULAR ANALYSISType: Extensible Order: Significant

NL Rel with Parent VT Concept Name VM Req Type Condition Value Set Constraint
1 CONTAINER EV (122292, DCM, "Quantitative Ventriculography Report”) 1 M
2 > HAS CONCEPT MOD INCLUDE DTID (1204) Language Of Content Item And Descendants 1 M
3 > HAS OBS CONTEXT INCLUDE DTID (1002) Observer Context 1-n M
4 > CONTAINS INCLUDE DTID (3602) Cardiovascular Patient Characteristics 1 U
5 > CONTAINS CONTAINER EV (122144, DCM, “Quantitative Analysis”) 1-n M
6 >> HAS OBS CONTEXT INCLUDE DTID (1002) Observer Context 1-n U
7 >> HAS OBS CONTEXT TEXT EV (111001, DCM, “Algorithm Name”) 1 M
8 >> HAS OBS CONTEXT TEXT EV (111003, DCM, “Algorithm Version”) 1 M
9 >> HAS OBS CONTEXT TEXT EV (122405, DCM, “Algorithm Manufacturer”) 1 M
10 >> CONTAINS IMAGE EV (121112, DCM, “Source of Measurements”) 1-n M
11 >>> HAS CONCEPT MOD CODE EV (G-A60B), SRT, “Cardiac Phase” 1 M DCID (12233) Cardiac Phase
12 >>> HAS CONCEPT MOD CODE EV (111031, DCM, “Image View”) 1 MC If Biplane Analysis DCID (3466) Plane Identification
13 >> HAS ACQ CONTEXT INCLUDE DTID (3205) Calibration 1-2 U VM = 1: Single plane analysis, VM = 2: Biplane analysis $CalibrationPlane = DCID (3466) Plane Identification
14 >> HAS ACQ CONTEXT INCLUDE DTID (3520) Hemodynamic Clinical Context 1 U
15 >> CONTAINS INCLUDE DTID (3206) VA Main Results 1 M
16 >> CONTAINS INCLUDE DTID (3207) AA Main Results 1 U
17 >> CONTAINS INCLUDE DTID (3208) Frame to Frame Results 1 U
18 >> CONTAINS INCLUDE DTID (3209) Centerline Wall Motion 1-2 U VM = 1: Single plane analysis, VM = 2: Biplane analysis
19 >> CONTAINS INCLUDE DTID (3210) Radial Based Wall Motion 1-2 U VM = 1: Single plane analysis, VM = 2: Biplane analysis
20 >> CONTAINS INCLUDE DTID (3211) Landmark Based Wall Motion 1-2 U VM = 1: Single plane analysis, VM = 2: Biplane analysis

Content Item Descriptions

Row 7 Identifies the Ventricular Analysis program
Row 8 Identifies the Ventricular Analysis program version
Row 9 Identifies the Ventricular Analysis program manufacturer
Row 10 Identifies the ES and ED images on which the analysis is based, for frame by frame analysis the analyzed image are specified in the frame by frame results (3208) template