TID 3107 Device Used

The Device Used Template allows recording of the use of interventional diagnostic and therapeutic devices.

The identification of one device used to deploy another device (e.g., a balloon catheter to deploy a stent) may be described with two entries, with one identified as a deployment device in the Concept Modifier of Row 6 of this template, and linked by the same Procedure Action ID in the Log Entry Qualifiers of the included TID 3010.

TID 3107Device UsedType: Extensible Order: Significant

NL Relation with Parent Value Type Concept Name VM Req Type Condition Value Set Constraint
1 CODE DCID (3422) Device Use Actions 1 M BCID (3429) Catheterization Devices
2 > HAS PROPERTIES CODE EV (121150, DCM, "Device Code") 1-n U Vendor or local bar coded values
3 > HAS PROPERTIES TEXT EV (121145, DCM, "Description of Material") 1 U
4 > HAS PROPERTIES NUM DCID (3423) Numeric Device Characteristics 1-n U
5 > HAS PROPERTIES CODE EV (G-C0E9, SRT, “Procedure site”) 1 U BCID (3630) Cardiovascular Anatomic Locations
6 > HAS CONCEPT MOD CODE EV (G-C0E8, SRT, “Has Intent") 1 U DT (121155, DCM, “Deployment")
7 > INCLUDE DTID (3010) Log Entry Qualifiers 1 U