C.3.1.1 Ultrasound Single and Multi-frame Pixel Formats Supported

The STD-US application profile requires that all ultrasound image objects only be stored using the values described in PS 3.3 US Image Module and the specializations used for the Ultrasound Single and Multi-Frame IODs.

In the role of FS-Updater or FS-Creator the application can choose any of the supported Photometric Interpretations described in PS 3.3 US Image Module to create an IOD. In the role of FS-Reader, an application shall support all Photometric Interpretations described in PS 3.3 US Image Module.

Table C.3-2 describes restrictions on the use of various Transfer Syntaxes with the supported Photometric Interpretations for both single and multi-frame images.

Table C.3-2DEFINED PHOTOMETRIC INTERPRETATION AND TRANSFER SYNTAX PAIRS

Photometric Interpretation Value Transfer Syntax Transfer Syntax UID
MONOCHROME2 Uncompressed RLE Lossless Image Compression 1.2.840.10008.1.2.1 1.2.840.10008.1.2.5
RGB Uncompressed RLE Lossless Image Compression 1.2.840.10008.1.2.1 1.2.840.10008.1.2.5
PALETTE COLOR Uncompressed RLE Lossless Image Compression 1.2.840.10008.1.2.1 1.2.840.10008.1.2.5
YBR_FULL RLE Lossless Image Compression 1.2.840.10008.1.2.5
YBR_FULL_422 Uncompressed JPEG Lossy 1.2.840.10008.1.2.1 1.2.840.10008.1.2.4.50
YBR_PARTIAL_422 Uncompressed JPEG Lossy 1.2.840.10008.1.2.1 1.2.840.10008.1.2.4.50